| Brand Name: | LeoTech |
| Model Number: | M1 Pro |
| MOQ: | 1 |
| Payment Terms: | T/T |
The M1 Pro sintering furnace is designed for high quality dental zirconia sintering. It contains 200 built-in editable circles with maximum temperature of 1550 degree.The M1 pro empowers advanced heating technology and accurate temperature control system, which meet your strict sintering demands of various brands of zirconia materials.
| Rated Frequency | 50Hz | Maximum Working Temperature | 1550°C |
| Rated Power | 2500W | Heating Rate | 1-50°C |
| Net Weight | 44.35kg | Gross Weight | 66.45kg |
| Dimentioins (cm) | 51*42*80 (cm) | Package Dimensions (cm) | 63*55*95 (cm) |
![]()
Imported brand silicon carbide rods are used, featuring uniform heating, corrosion resistance, high temperature resistance, and simple installation and maintenance.
![]()
It personalized large LCLD touch screen, friendly interface, one-key start.
| Code | Error reasons | Solutions |
|---|---|---|
| 1 | Temperature sensor failure | Replace temperature sensor orconnecting cable |
| 2 | The polarity of the temperature sensor connect incorrectly | Adjust the temperature sensor connect |
| 3 | Heater failure | Check the lines or replace |
| 4 | The chamber of furnace seal is poor, too large a temperature leakage and the heating ability is weakened | Replace the furnace chamber and check if there's any oxidation in the connection of the heating elements |
| 5 | Unstable voltage | Check the wiring and install voltage regulator |
| 6 | Tray lifting failure | Check the lift limit switch or drive motor |
| 7 | Tray descending failure | Check the descend limit switch or drive motor |
| 8 | Abnormal communication of thermostat | Check thermostat or wiring |
| 9 | Power failure warning | Power up and operate accordingly |
![]()
![]()